8-4 Measurement 
  Abdomen measurements - Used for measurements of abdominal organs (liver, gall 
bladder, pancreas and kidney, etc.) and large abdominal vessels.   
  OB measurements - Used for measurements of fetal growth indexes (including EFW) as 
well as GA and EDD calculations. The fetus can be evaluated through growth graph 
analysis and fetal biophysical profile.   
  Cardiac measurements – Used for left ventricle function measurements and 
measurements of main artery and vein parameters, etc. 
  Gynecology measurements - Used for the uterus, ovary and follicles, etc. 
  Small Part measurements – Used for small parts such as thyroid. 
  Urology measurements - Used for prostate, seminal vesicle, renal, adrenal, micturated 
and testicle volume. 
  Orthopedics measurements – Used for hip joint measurement. 
  Vascular measurements – Used for carotid, cerebral, upper and lower extremities vessels, 
etc.  
  EM measurements – Used for EM related full functional measurement. 
8.4 Measurement Accuracy 
Table 1 Error of 2D Images 
Parameter   Value Range  Error 
Distance  Full screen  Within ±2% 
Area  Full screen  Within ±6% 
Circ  Full screen  Within ±7% 
Angle  Full screen  Within ±3%. 
Volume    Full screen  Within ±7 % 
Table 2 iScape Measurements 
Parameter   Probe Type  Error 
Distance 
Full screen  Within ±5 % (L14-6NE, L12-3E, P4-2, D7-
2E, 3C5A, 7L4A probe) 
Full screen  Within ±10 % (V11-3, CB10-4E probe) 
Table 3 Time/Motion Measurements 
Parameter   Value Range  Error  
Distance    Full screen  Within ±3 % 
Time    Timeline Display  Within ±2% 
Heart rate  Timeline Display    Within ±3% 
Velocity (PW 
mode) 
10-200cm/s: (for 
L14-6NE, L12-3E, 
CB10-4E, 7L4A) 
10-300cm/s: (for 
P4-2, D7-2E, V11-3, 
3C5A) 
 
When angle≤60º, ≤5%.