DC7 Diagnostic Ultrasound System
21
Figure 14 Send Image to DICOM Storage
2. In Storage Preset -> DICOM Service, set the default Storage server (one or
more).
3. If <Save> on control panel is preset as saving single-frame image, the
system will save and send a single-frame image to remote DICOM storage
when <Save> is pressed. If <Save> on control panel is preset as saving
multi-frame image, the system will save and send a multi-frame image to
remote DICOM storage when <Save> is pressed.
C. Send to DICOM Storage by user-defined shortcut key
1. In System Preset -> Key Config, the NULL key can be set as “Send Image to
DICOM storage”, as shown in Figure 15.