6 - 46 Operator’s Manual
  6  Image Acquisition
No. Device 
Name
Port Description
1 Auxiliary 
output 
power 
supply 
cable
Power supply port Plug the auxiliary output power supply cable to the port of 
the magnetic navigator, and connect the power supply.
NOTE:
Please use the auxiliary output power supply cable 
equipped with the Ultrasound System.
2USB data 
cable
USB port Plug one end of the USB data cable to the USB port of the 
magnetic navigator, and another end to the USB port of the 
Ultrasound System.
3 Magnetic 
generator
Transmitter port Plug the magnetic generator to the Transmitter port. Place it 
around the patient’s shoulder or two sides of the patient’s 
abdomen. The magnetic generator should be frontal-faced 
with the lesion. The frontal face of the magnetic generator 
keeps 15-60 cm from the target within the scope 30 cm up/
down and towards probe.
NOTE:
Keep the magnetic generator away from the metal 
devices or electromagnetic interference.
4-9 / Positioning 
(probe) sensor port
Motion (abdomen) 
sensor port
 1. Plug two sensors to positioning (probe) sensor port
  , and motion (abdomen) sensor port   
 respectively.