EasyManuals Logo

Mindray Resona 7 Operator's Manual

Mindray Resona 7
492 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #443 background imageLoading...
Page #443 background image
F Electrical Safety Inspection
Operator’s Manual F - 3
• normal polarity with open neutral (Single Fault Condition).
• reverse polarity with open neutral (Single Fault Condition).
LIMITS
• For UL 60601-1:
–300 μA in Normal Condition.
– 1000 μA in Single Fault Condition.
• For IEC 60601-1:
–500 μA in Normal Condition.
– 1000 μA in Single Fault Condition.
F.6 Enclosure Leakage Test
The following outlet conditions apply when performing the Enclosure Leakage test.
• normal polarity (Normal Condition).
• reverse polarity (Normal Condition).
• normal polarity with open neutral (Single Fault Condition).
• reverse polarity with open neutral (Single Fault Condition).
• normal polarity with open earth (Single Fault Condition).
• reverse polarity with open earth (Single Fault Condition).
LIMITS
• For UL 60601-1
–100 μA in Normal Condition.
–300 μA in Single Fault Condition.
• For IEC 60601-1:
–100 μA in Normal Condition.
–500 μA in Single Fault Condition.
F.7 Patient Leakage Current
Patient leakage currents are measured between a selected applied part and mains earth. All
measurements have a true RMS only.
The following outlet conditions apply when performing the Patient Leakage Current test.
• normal polarity (Normal Condition).
• reverse polarity (Normal Condition).
• normal polarity with open neutral (Single Fault Condition).
• reverse polarity with open neutral (Single Fault Condition).
• normal polarity with open earth (Single Fault Condition).
• reverse polarity with open earth (Single Fault Condition).
LIMITS
For BF applied parts:
•100 μA in Normal Condition.
•500 μA in Single Fault Condition.

Table of Contents

Other manuals for Mindray Resona 7

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Mindray Resona 7 and is the answer not in the manual?

Mindray Resona 7 Specifications

General IconGeneral
BrandMindray
ModelResona 7
CategoryMedical Equipment
LanguageEnglish

Related product manuals