17 Probes and Biopsy
Operator’s Manual 17 - 53
NGB-034
Perform the following procedure:
1. Install the needle-guided bracket:
a. Put on the sterile probe sheath.
b. Select a proper needle-guided bracket, and match the locating groove with the tab of the
probe. Mount the bracket onto the probe.
c. Tighten the pinch nut of the needle-guided bracket to confirm that the needle-guided
bracket is properly installed on the probe.
2. Adjust the needle angle to the proper shift as required:
a. Loosen the angle pinch nut.
b. Adjust the angle block to the desired level.
c. Tighten the angle pinch nut.
1 Needle fixing nut 2 Guiding groove
3 Angle fixing nut 4 Angle adjusting base
5 Needle type dial scale 6 Needle adjusting base
7 V-shape cover 8 Lock pin
9 Needle guide hole 10 V-shaped guiding block
11 Clamp
1
3
4
2
5
6
7891011