UA-66 Urine Analyzer Operator’s Manual 
- 12 - 
3  Setup 
At the main menu( See Figure 2.5 ), use  ▲  and  ▼ as necessary to select    “Options” , and 
press START to enter “Options” menu: 
Options 
Strip  Printer 
Critical Value  Serial Port 
Grade  Serial No. 
Realtime Clock 
Figure 3.1 
Using this menu, the operator can: 
♦ Correct the system clock; 
♦ Set the options for printer and serial port; 
♦ Select the reagent strip , testing results output unit and critical values; 
♦ Show serial No. of the instrument. 
3.1  Reagent Strip 
NOTE: The name of the reagent strip being used must be the same as the reagent strip 
selected by the “Options>Strip” menu. Use of any other strip will cause erroneous results. 
At the “Options” menu( See Figure 3.1 ), use ▲ and ▼  to select “Strip”, press START to enter 
the “Options>Strip” menu: 
 
10 Para. 
11 Para. 
 
Figure 3.2 
Use ▲  and ▼ as necessary to select the designated reagent strip, press START to validate the 
new selection. 
3.2  Critical Value 
Critical values are a set of values to be compared with the testing results. If the result of a 
certain parameter is the same as or higher than the corresponding critical value, a “*” will be 
inserted in front of the corresponding row while the report being displayed and printed. 
At  the  “Options”  menu,  use 
▲  and  ▼   to select  “Critical  Value”,  press  START  to  enter  the