UA-66 Urine Analyzer Operator’s Manual
- 12 -
3 Setup
At the main menu( See Figure 2.5 ), use and as necessary to select “Options” , and
press START to enter “Options” menu:
Options
Strip Printer
Critical Value Serial Port
Grade Serial No.
Realtime Clock
Figure 3.1
Using this menu, the operator can:
Correct the system clock;
Set the options for printer and serial port;
Select the reagent strip , testing results output unit and critical values;
Show serial No. of the instrument.
3.1 Reagent Strip
NOTE: The name of the reagent strip being used must be the same as the reagent strip
selected by the “Options>Strip” menu. Use of any other strip will cause erroneous results.
At the “Options” menu( See Figure 3.1 ), use and to select “Strip”, press START to enter
the “Options>Strip” menu:
Options>Strip
10 Para.
11 Para.
Figure 3.2
Use and as necessary to select the designated reagent strip, press START to validate the
new selection.
3.2 Critical Value
Critical values are a set of values to be compared with the testing results. If the result of a
certain parameter is the same as or higher than the corresponding critical value, a “*” will be
inserted in front of the corresponding row while the report being displayed and printed.
At the “Options” menu, use
and to select “Critical Value”, press START to enter the