UA-66 Urine Analyzer Operator’s Manual
- 30 -
6. Remove the feed table insert and rinse thoroughly with distilled water to remove any residual
solution.
7. Wipe the feed table insert in dry with an absorbent tissue.
8. Replace the feed table insert in the feed table, as instructed in 7.2 Daily Cleaning.
7.4 Periodic Cleaning
If the feed table insert becomes excessively dirty and cannot be adequately cleaned as
d
escribed in 7.2 Daily Cleaning, the following procedure should be used:
1. Remove the feed table insert from the table, as described in 7.2 Daily Cleaning.
2. Wipe the strip holding channel of the feed table insert with 0.1N sodium hyndroxide (NaOH)
using a cotton-tipped applicator stick or other suitable absorbent material. Thoroughly swab
the channel area until it is clean.
NOTE: Do not wipe the white plastic calibration chip with the NaOH. Do not use any
material that will scratch the feed table insert. Do not allow the NaOH to come into
contact with the white plastic calibration chip or with any other portion of the feed table
or instrument.
3. Rinse the feed table insert channel thoroughly with water to remove any residual NaOH.
4. Wipe the feed table surface and feet table mounting groove with a damp cloth. Wipe the feed
table dry with an absorbent tissue.
5. Replace the feed table insert in the feed table, as instructed in 7.2 Daily Cleaning.