UA-66 Urine Analyzer Operator’s Manual 
-19- 
5  samples  are  listed  on  the  displayer  and  the  sample  No.  next  to  the  latest  sample  is 
automatically selected. Sample No. and the time when being tested are used to identify a sample. 
A prefixed “*” indicates that the sample has at least a result higher than the corresponding critical 
value, and a prefixed “ ” indicates a sample marked by the operator during the testing. “--:--:--” 
indicates no testing was done under this No. Press ▲ and ▼ to change the selection. The list will 
scroll when operator try to select the sample previous to the sample listed on the top or next to the 
sample listed on the bottom. The operator can also enter the desired sample No. directly by 
pressing FUN to pop up an input box: 
Sample No. 
013 
Figure 4.3 
Press  ►  to change the select digit and press  ▲ /  ▲ to modify. Press  START to validate the 
input.   
When the sample No. selection is done, press START to start testing under the mode selected 
before. 
4.3  The Single Mode 
012 
013 
LEU    ------------------- 
NIT    ------------------- 
URO    ------------------- 
PRO    ------------------- 
PH      ------------------- 
BLD    ------------------- 
SG      ------------------- 
KET    ------------------- 
BIL    ------------------- 
GLU    -------------------
Press 
’START’
Figure 4.4 
Figure 4.1 shows the display of single mode testing. The display area is divided into 4 blocks. 
The top-left is the sample number area, the bottom-left is the report area, the top-right is the 
progress area, and the bottom-right is the user prompt area.   
Press START to start testing. The internal timer of the instrument begins to count down and a 
progress bar is shown under the sample number at the progress area. With 3 short noises from 
the buzzer, and the message “Dip 012” shown at the prompt area, the operator is prompted with 
immerse all reagent areas of the reagent strip in  the urine; following the long noise from the 
buzzer and a “Place 012” prompt, place the reagent strip on the feed table. Then the feed table 
will move to the inside of the instrument. The instrument will test the stripe block by block when 
the feed table moving outside. When the feed table return to the original position the entire test 
cycle is over, and the result of the test is displayed at the report area, and output via the thermo 
printer. 
Press 
START to start a new test cycle.