UA-66 Urine Analyzer Operator’s Manual
-19-
5 samples are listed on the displayer and the sample No. next to the latest sample is
automatically selected. Sample No. and the time when being tested are used to identify a sample.
A prefixed “*” indicates that the sample has at least a result higher than the corresponding critical
value, and a prefixed indicates a sample marked by the operator during the testing. “--:--:--”
indicates no testing was done under this No. Press and to change the selection. The list will
scroll when operator try to select the sample previous to the sample listed on the top or next to the
sample listed on the bottom. The operator can also enter the desired sample No. directly by
pressing FUN to pop up an input box:
Sample No.
013
Figure 4.3
Press to change the select digit and press / to modify. Press START to validate the
input.
When the sample No. selection is done, press START to start testing under the mode selected
before.
4.3 The Single Mode
012
013
LEU -------------------
NIT -------------------
URO -------------------
PRO -------------------
PH -------------------
BLD -------------------
SG -------------------
KET -------------------
BIL -------------------
GLU -------------------
Press
’START’
Figure 4.4
Figure 4.1 shows the display of single mode testing. The display area is divided into 4 blocks.
The top-left is the sample number area, the bottom-left is the report area, the top-right is the
progress area, and the bottom-right is the user prompt area.
Press START to start testing. The internal timer of the instrument begins to count down and a
progress bar is shown under the sample number at the progress area. With 3 short noises from
the buzzer, and the message “Dip 012” shown at the prompt area, the operator is prompted with
immerse all reagent areas of the reagent strip in the urine; following the long noise from the
buzzer and a “Place 012” prompt, place the reagent strip on the feed table. Then the feed table
will move to the inside of the instrument. The instrument will test the stripe block by block when
the feed table moving outside. When the feed table return to the original position the entire test
cycle is over, and the result of the test is displayed at the report area, and output via the thermo
printer.
Press
START to start a new test cycle.