Set A Non-Invasive Blood Pressure Measurement Interval
The Accutorr 7 monitor allows the clinician to set a pre-configured interval for a
NIBP measurement to be obtained automatically.
To set an automatic interval on the Accutorr 7 monitor:
a. Select NIBP parameter area to access the NIBP Setup menu.
b. Set Interval.
• Interval can be set to 1 min, 2 min, 2.5 min, 3 min, 5 min, 10 min, 15
min, 20 min, 30 min, 1 h, 1.5 h, 2h: The monitor automatically measures
NIBP based on the specified time interval.
• Interval can be set to Program 1/Program 2: The monitor automatically
measures NIBP based on a program customized by the user.
c. Select the Main quick key or to return to the main display.
*If NIBP Interval is enabled in the Remote Control menu of BeneVision Central
Monitoring System (BeneVision CMS): Selecting Interval on the BeneVision
CentralStation and BeneVision WorkStation allows the clinician to set an
automatic interval on the Accutorr 7 monitor.
Configure A Custom Program
The Accutorr 7 monitor allows the clinician to perform NIBP measurement based
on a customized program.
To configure a custom program on the Accutorr 7 monitor:
a. Select NIBP parameter area to access the NIBP Setup menu.
b. Select Custom Program >>.
c. Set Program to Program 1 or Program 2.
d. Set Duration and Interval for each measurement cycle.
e. Select Ok to confirm.
f. Set Interval to Program 1 or Program 2.