15
NOTE: The mode can be selected only by using the CSW-31
software.
5.4.4.1 Fixed Dose Rate Alarm Level:
There are 8 different levels for dose rate
alarm:
disabled, 10, 50, 100, 500, 1 000, 5
000, 10 000 or 50 000 µSv/h or mrem/h.
The user may select any one of these to be the active alarm
threshold level.
Note: All these dose rate alarm threshold levels can be changed and stored
into the instrument’s internal memory with the CSW-31 software.
To see the current dose rate alarm threshold level:
Activate the menu (Short [Ξ]) and scroll (Short [Ξ]’s) until
display changes to FX.RAL
The current dose rate alarm threshold value is displayed on
right side of the display.
To change the current dose rate alarm level:
While the current dose rate alarm level is displayed, press
[
]shortly; the next level in row starts blinking
Scroll with [Ξ]’s shortly until the desired value is given
Press an intermediate [
]
The new alarm limit is activated
FX.RALSUBmenu
...
FX.RAL
OFF
10.00µSv/h
50.00µSv/h
Choose
Option
100.0µSv/h
500.0µSv/h
1.000mSv/h
5.000mSv/h
10.00mSv/h
50.00mSv/h
Allnumericalvaluesofthelistcanbechanged
usingtheCSW31software.
...