22
SQ.DALSUB‐menu
...
↓
↑
SQ.DAL
500.0mSv
↓
↑
OFF*)
↓
↑
100.0µSv*)
→
Choose
Option
↓
↑
500.0µSv
↓
↑
1.000mSv
↓
↑
5.000mSv
10.00mSv
50.00mSv
100.0mSv
*)OnlyOFForlowestvalueonthe
listcanbechosen,whentheSequentialalarm
modeisactivated.Othervaluesofthelistcan
beviewedonly.
↓
↑
...
Note: The alarm condition can be reset only by resetting the cumulative
dose
.
5.4.7 Time to Dose Alarm (TM.2.AL)
NOTE: the Time to Dose Alarm function can
be enabled/disabled with the CSW-31
Software only!
The Time to Dose Alarm function can be enabled to show the
estimate of the remaining work time until the next dose alarm is
activated in the current dose rate field.
In the Fixed dose alarm mode, the procedure how to show and
change the valid dose alarm level is described earlier and will set
a new reference point for this calculation.
In the Sequential dose alarm mode the Time to Dose Alarm
function is not in use but the display shows 00h 00m.
In case the dose alarm system is set to OFF state the function is
disabled, indicated by the DISAB message on display’s right side