40
menu
Configuration 2: The Audible chirp is disabled (Visual chirp is
*)
X)
Configuration 3: The Audible chirp is enabled (Visual chirp is
*)
X)
Configuration 4: The Visual chirp is disabled (Audible chirp is
*)
X)
Configuration 5: The Visual chirp is enabled (Audible chirp is
*)
X)
Configuration 6: The pulse rate from the detector is divided by a
value from the fixed list:
1/1, 1/2, 1/5, 1/10, 1/20 and 1/50
*)
The state X is either Enable or Disable
6.4 Show and Change Dose Rate Alarm Level
(FX.RAL/SQ.RAL)
The Dose rate alarm is always measured from the internal
tube even when an external detector has been inserted.
This is to ensure the protection of the user of any telepole
device.
This is a configurable function with 3 options:
Option Effect
Configuration 1: Disabled from the menu.
Configuration 2: The current alarm level can be displayed but
cannot be changed using the instrument’s push
button.
Configuration 3: The Alarm Level is displayed and can be changed
using the push button.
6.5 Show and Change External Dose Rate / Count Rate
Alarm Level (XT.RAL/XT.CRA)
The Dose rate and Count rate alarm is measured from the
external detector.
This is a configurable function with 3 options:
Option Effect
Configuration 1: Disabled from the menu.
Configuration 2: The current alarm level can be displayed but
cannot be changed using the instrument’s push
button.
Configuration 3: The Alarm Level is displayed and can be changed
using the push button.
6.6 Show and Change Dose Alarm Level (FX.DAL/SQ.DAL)
The Dose alarm is always measured from the internal
tube even when an external detector has been inserted.
This is to ensure the protection of the user of any telepole
device.