EasyManua.ls Logo

Mirion Technologies RDS-31 S - Other Displayed Messages; Low Battery Warning; Low Battery Alarm; Dose Rate Alarm (Blinking Display)

Mirion Technologies RDS-31 S
54 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
37
Connection The menu item of CONN,
indicating communication states. The normal
button operation is allowed.
Histogram The menu item of HISTO,
indicating histogram function state. The
normal button operation is allowed.
Chg. hist.
state
The submenu of HISTO, allowing
configuration of histogram functions.
Manual histog Takes manual histogram sample.
Advanced The menu item of Advanced. The normal
button operation is allowed.
5.6 Other Displayed Messages
5.6.1 Low Battery Warning
An audible low battery warning signal combined with a blinking
battery symbol indication repeated every 6 minutes when the
battery capacity is 10% < Capacity < 30%
Low battery warning:
_______ _ _______
(Long – Short – Long
Bleep in five minutes intervals). The dose rate measurement
continues normally.
5.6.2 Low Battery Alarm
The display shows "LOBAT" blinking and the audible alarm is
activated when the battery capacity is <10%.
Low battery alarm:
_______ _ _______
(Long – Short – Long Bleep
in 20 second intervals).
Note: The dose rate measurement is prevented.
5.6.3 Dose Rate Alarm (Blinking Display)
The dose rate display is blinking and the continuous audible
alarm is on. The audible alarm can be reset by pressing the push
button.
Dose rate alarm: ------ (two beeps per second).
5.6.4 Dose Alarm
The dose rate display is on and the current dose alarm level
alternates in the display and the continuous audible alarm is on.
The audible alarm can be reset by pressing the push button.
Dose alarm: - - - - - - (one beep per two seconds).

Table of Contents