23
To check the status of the Time to Dose Alarm:
Activate the menu (Short [Ξ]) and scroll (Short [Ξ]’s) until
display changes to TM.2.AL
The current status is displayed (the remaining time to
reach the current dose alarm level in form of xxh xxm or
disabled = DISAB).
5.4.8 Diagnostics (DIAG)
The Diagnostics function of the RDS-31 is
used to manually activate the internal
checking functions of the high voltage, the
operating voltage and the battery capacity. During this process
the user may also check the operation of all the display pixels,
the visual alarm LED, the audible alarm and the vibration
alarm. The instrument software revision, XX and XXX can also
be seen. The Diagnostics sub-functions are provided in the
following table
Display Function | Description
8.8.8.8.8
*)
Main test functions can be activated under this
All pixels are turned on
The display shows different symbol configurations.
The Buzzer is activated
The Vibration motor is activated
The RED alarming LED is activated
The measured performance parameters are stored into the
Histogram memory for later diagnostics
V. 2.01.1 Current revision of the firmware (Major. Minor)
CAL.CHECK Start the calibration check procedure
Chooses the target dose rate that is used. After a fixed 1
minute waiting time the instrument measures a dose of 100
µSv (time calculated from the rate). In case the measured
dose is within 10% from expected dose OK, otherwise warn.
B:74% Battery Capacity (here 74 % of the full value)
LOADER Activate Firmware loading / Restart instrument
*)
Underlined and italic cursive is in use when the display is blinking thus
indicating an immediate action followed with an intermediate pressing of
the [
] button.