24
To activate the diagnostics functions:
Activate the menu (Short [Ξ]) and scroll (Short [Ξ]’s) until
display changes to DIAG
Press [
]shortly
The display shows blinking “8.8.8.8.8”
DIAGSUB‐menu
...
↓ ↑
DIAG
B:65%
InformationField
Lastmeasuredbatterycapacity
↓ ↑
LOADER
Command
StartFWupdate/Resetmicrocontroller
↓ ↑
8.8.8.8.
→
ChooseOption
StartDiagnostics
↓ ↑
V2.04.1
InformationField
CurrentFW‐
revision
↓ ↑
CAL.CHEC
K
Command
StartCalibrationCheck
↓ ↑
...
To perform the diagnostics:
While the “8.8.8.8.8” is blinking
Activate diagnostics with an intermediate [
]
All display pixels are turned on
The display shows different symbol configurations.
The buzzer is activated
The vibration motor is activated
The RED alarming LED is activated
The display indicates the capacity of the battery B:xx (%)
Note: Do not press any buttons while the meter is performing diagnostics,
otherwise the function will be interrupted. If diagnostics is interrupted
perform a new diagnostics session.
To perform calibration:
While the “8.8.8.8.8” is blinking
Press [Ξ] shortly
The display provides the Firmware revision
While the Firmware revision is displayed
Press [Ξ] shortly