14
4.8 Analogue I/O Connector
A standard 15-way D-type socket used for interfacing with external equipment.
This may be already in use depending on the type of RGA system purchased.
This interface offers the following functionality:
Two analog outputs 0 to +10V 12bit
Four, quasi-differential analog inputs, -11 to +11V, with a maximum voltage on the return of ±0.5V
±15V power outputs both fused at 120mA, fuses are self resetting.
NOTE:
The 120mA maximum load for the power supply outputs is shared between the Digital and Analogue
connectors.
Pin Function Description
1 -15V -15V supply output, fused at 120mA
2 Analogue Input 4 Return
3 Analogue Input 3 Return
4 Analogue Input 2 Return
5 Analogue Input 1 ±11v
6 Composite Return Differential Output Return
7 Electrometer Differential Output Direct Electrometer output from 50Meg feedback
resistor
8 0V
9 +15V +15V supply output, fused at 120mA
10 Analogue Input 4 ±11v
11 Analogue Input 3 ±11v
12 Analogue Input 2 ±11v
13 Analogue Input 1 Return
14 Analogue Output 1 0 – 10v
15 Analogue Output 2 0 – 10v