EasyManua.ls Logo

MKS Microvision2 - 4.8 Analogue I;O Connector

MKS Microvision2
50 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
MKS Instruments UK Ltd
Microvision2 Hardware Manual – SP101015.102 August 2010
14
4.8 Analogue I/O Connector
A standard 15-way D-type socket used for interfacing with external equipment.
This may be already in use depending on the type of RGA system purchased.
This interface offers the following functionality:
Two analog outputs 0 to +10V 12bit
Four, quasi-differential analog inputs, -11 to +11V, with a maximum voltage on the return of ±0.5V
±15V power outputs both fused at 120mA, fuses are self resetting.
NOTE:
The 120mA maximum load for the power supply outputs is shared between the Digital and Analogue
connectors.
Pin Function Description
1 -15V -15V supply output, fused at 120mA
2 Analogue Input 4 Return
3 Analogue Input 3 Return
4 Analogue Input 2 Return
5 Analogue Input 1 ±11v
6 Composite Return Differential Output Return
7 Electrometer Differential Output Direct Electrometer output from 50Meg feedback
resistor
8 0V
9 +15V +15V supply output, fused at 120mA
10 Analogue Input 4 ±11v
11 Analogue Input 3 ±11v
12 Analogue Input 2 ±11v
13 Analogue Input 1 Return
14 Analogue Output 1 0 – 10v
15 Analogue Output 2 0 – 10v

Related product manuals