203 Slv Internal error Slv 203 Hard Fault of fuzzy regulator of RH (this
should never occur)
Switch the device on and off using the
network switch.
Contact authorised service in case
that the problem persists.
400 Comm Internal error Com 400 Soft Fault of reading EEPROM Com Switch the device on and off using the
network switch.
Contact authorised service in case
that the problem persists.
401 Comm Internal error Com 401 Soft Fault of writing to EEPROM Com Switch the device on and off using the
network switch.
Contact authorised service in case
that the problem persists.
403 Comm Internal error Com 403 Soft Unexpected reset Com Switch the device on and off using the
network switch.
Contact authorised service in case
that the problem persists.
10 DEVICE MAINTENANCE
10.1 MAINTENANCE PLAN
In case of long-term cut off of the device discharge water from the cask, generator and waste vessel. Avoid any device
pollution with ditch water. Dry the device and its parts and we also recommend its washing with suitable disinfection means.
Perform any replacement, setting, cleaning, etc. only with cooled down and disconnected device!
In case of the device to be used in any other but recommended way, the device protection may be impaired.
Frequency Controlled part Performed by Note
continuously Control of hoses and hose connections for leakage,
cleanness of mesh on submersible pump
staff
continuously Control of water quantity and cleanness in the cask
(min. level height 25 cm)
staff Pollution may occur after long-term
set-off
continuously Control of waste vessel cleanness staff Pollution may occur after long-term
set-off
continuously Cleaning of inner sheets, chamber, shell, touch
screen
staff As needed
continuously Control of door tightness staff Replacement, as needed
min. 1/2 year Visual control of integrity of insulation of fl exible
temperature sensors
staff
1 year Control of door locking mechanism service
1 year Condensate cleaning in cooling circuit service Dirt between lamellas impairs the cooling
circuit function
1 year Protective thermostat control service
1 year Relief valve functionality control service
1 year Control of electric installation – mainly integrity
of insulation of conductors, intake, connection
and protective clamps, strength of conductors
connection in clamps
service Protective connection resistance is
< 0,1 Ohm, the intake resistance is not
included
1 year Control of hose connection CO
2
service
Instructions for use
CLIMACELL EVO_np_en_1401_mmm_V1.01 41