11 TECHNICAL DATA II
11.1 OPERATION DATA
Type 111 / 111-c 222 / 222-c 404 / 404-c 707 / 707-c 1212 / 1212-c
Operation temperature *) from 0 °C to °C 100 (decontamination 160 °C) xx
from -20 °C to °C 100 (decontamination 160 °C) 70
Temperature precision space at 10 °C
at 37 °C
approximately
(±)°C
< 0,5 < 0,5 < 1 < 1 <1,5
< 0,5 < 0,5 < 1 < 1 <1,5
time approximately
(±)°C
< 0,2 < 0,2 < 0,3 < 0,4 < 0,8
Time of heating to 37 °C from ambient
temperature
min <11 <11 <22 <13 <30
Time of cooling from temperature
22 °C to 10 °C
min < 21 / < 11 <17 / <14 <19 / <21 <21 / <22 <21
Time of restoration after
30 s door opening according
to DIN 12880
at 37 °C min < 5 < 2 < 10 6 10
at 50 °C min < 6 < 3 < 13 6 10
Relative humidity range % 10 - 95
Precision
T
chamber
21 °C
time
approximately
(±) % RH
< 2
Concentration of CO
2
% 0,1 - 20 xx
Required pressure of CO
2
bar/psi 0,3 - 0,7 / 5 - 10 xx
All the data apply in empty chamber (without samples on screens) for ambient temperature of 20 - 22 °C, 100 % fan revolutions,
feeding voltage 230 V ± 10 %.
*) Option of cooling up to -20 °C, performance with decontamination – temperature 160 °C. If there is used lighting or CO
2
sensor, the allowed temperature is lower.