EXACTA
11
INTRODUCING THE AUTOCLAVE
– HEATING
Heating of the water up to boiling point
– TERMODYNAMIC VACUUM 2
Second air bleeder by thermodynamic
system
– HIGH PRESSURE
Water vaporization and chamber high
pressure
– PROCESS
Exposure of the material for the select-
ed temperature and time
– DRY (or DECOMP in case of liquids)
Vacuum drying of the material
– LEVEL
Leveling and forced ventilation with
sterile air
– END
End of the cycle (open the door and re-
covery of the material)
MAIN FEATURES
– Patented initial vacuum system
– Very fast cycles
– Simple cycle selection and pro-
gramming through front panel push
buttons
– safety handle to prevent the opening
of the door during the sterilization cy-
cle; the door can be opened only
when the inside pressure of the
chamber is lower than 0,60 bar;
– safety valve for the steam discharge
in case of breakdown of the controls
and over pressure >2,4 bar (TÜV certi-
fication)
– alarm signaling for wrong operation
and on line self test;
– safety thermostat;
– EMI protection devices;
– separation of the used and distilled
water through different tanks;
– vacuum drying, as foreseen in the
hospital field;
– "Vacuum Test" to verify the seal of the
chamber and of the hydraulic circuit.
– TÜV GS certification.