EXACTA
11
INTRODUCING THE AUTOCLAVE
HEATING
Heating of the water up to boiling point
TERMODYNAMIC VACUUM 2
Second air bleeder by thermodynamic
system
HIGH PRESSURE
Water vaporization and chamber high
pressure
PROCESS
Exposure of the material for the select-
ed temperature and time
DRY (or DECOMP in case of liquids)
Vacuum drying of the material
LEVEL
Leveling and forced ventilation with
sterile air
END
End of the cycle (open the door and re-
covery of the material)
MAIN FEATURES
Patented initial vacuum system
Very fast cycles
Simple cycle selection and pro-
gramming through front panel push
buttons
safety handle to prevent the opening
of the door during the sterilization cy-
cle; the door can be opened only
when the inside pressure of the
chamber is lower than 0,60 bar;
safety valve for the steam discharge
in case of breakdown of the controls
and over pressure >2,4 bar (TÜV certi-
fication)
alarm signaling for wrong operation
and on line self test;
safety thermostat;
EMI protection devices;
separation of the used and distilled
water through different tanks;
vacuum drying, as foreseen in the
hospital field;
"Vacuum Test" to verify the seal of the
chamber and of the hydraulic circuit.
TÜV GS certification.