EXACTA
33
– START: If pressed with door open
and LED READY switched on, the cy-
cle will not start and the Timer display
will indicate the message F1 to signal-
ize the operator’s error (LED ALARM
switches on). Press any key to reset
this signaling and restore the opera-
tion. If pressed with door closed but
with LED READY switched off, the
program will not start. The timer dis-
play will show the alarm message F6
or F2.
– STOP; if pressed for more than 3 se-
conds during the cycle operation,
the program will be cleared. The tim-
er display will show the error message
F3 (LED ALARM switches on). The
pressure downs, and as reached a
value next to 0,00 bar, the Tempera-
ture display will show the message
End to signalize that the load may be
recovered. Press any key to reset this
signaling.
The selection may be carried out with door
open or close.
Pushing more times the Temperature and Load
keys you can choose 6 combinations of stand-
ard temperature/pressure/time values relating
to 6 sterilization cycles.
For each selection the corresponding LED will
switch on.
PARAMETERS OF THE STANDARD
STERILIZATIONCYCLES
CHOOSING A STANDARD STERILISATION PROGRAM