32
The display reports:
at open or closed door (before the
START command), the pressure (bar)
of the selected cycle and the pres-
sure in the chamber;
after the START command, the cur-
rent value of the chamber pressure
(bar); the value increases or de-
creases depending on the current
phase.
The display is used to signalize different func-
tions and status of the autoclave:
at open or close door and before the
START command, the selected sterili-
zation time (in minutes);
at close door and after the START
command, some messages show the
various phases of the cycle:
A flashing spot at the right of the display signal-
izes that the value decreases as the phase pro-
ceeds.
Two LEDs indicate the water level of the dis-
tilled water tank:
empty (first installation)
full (after the filling or a topping up)
normal (after some cycles)
topping up request (after more cy-
cles)
* THE START COMMAND IS DISABLED IF
MIN IS FLASHING
The used water tank is provided with maximum
level detector. As the tank is full, the Timer dis-
play shows hyphens as a demand for immedi-
ate emptying.
* WITH THIS SIGNALLING ON THE DISPLAY,
THE CYCLE WILL NOT START AND THE
TIMER WILL SHOW THE MESSAGE F6
(DEMAND FOR TANK EMPTYING).