EXACTA
46
* DO NOT LET THE CALCAREOUS OR DIRTY
RESIDUALS TO ACCUMULATE ON THE
GASKET, BECAUSE THEY CAN CAUSE ITS
DAMAGING OR BREAKAGE.
* DO NOT WASH THE AUTOCLAVE WITH DI-
RECT OR HIGH PRESSURE WATER JETS,
AS POSSIBLE INFILTRATION IN THE ELEC-
TRICAL CIRCUITS COULD COMPROMISE
THE REGULAR OPERATION OF THE APPA-
RATUS AND OF THE SAFETY CIRCUITS.
* NEVER USE DISINFECTANT PRODUCTS TO
CLEAN THE CHAMBER.
* CLEAN THE OVERALL DIAMETER OF THE
CHAMBER, AS RESIDUALS IN THE TOP
COULD COMPROMISE THE REGULAR OP-
ERATION OF THE ELECTROVALVES.
* THE BACTERIOLOGICAL FILTER CAN BE
MONTHLY MAINTAINED THROUGH A STER-
ILIZATION CYCLE AT 121°C (PRO-
GRAMME FOR WRAPPED MATERIAL).
GASKET AND PORT-LIGHT CLEANING
OUTER SURFACES CLEANING AND DISINFECTING
CHAMBER, HOLDER, AND TRAYS CLEANING
BACTERIOLOGICAL FILTER REPLACEMENT