EXACTA
34
* FOR SELECTING THE DISINFECTING CY-
CLE AT 105°C IT IS NECESSARY TO USE
THE SOFTWARE KEY (DRY OFF KEY).
The selection may be carried out with door
open or close.
– Press DRY-OFF key and more times
the Temperature key until LED 105°C
switches on. Release the keys;
– LED DIS switches on;
– TEMPERATURE display shows the value
105°C (for 4 seconds ) then switches
to the temperature value in the
chamber;
– PRESSURE display shows the process
pressure value of 0,20 bar (for 4 se-
conds) then switches to the current
value pressure in the chamber;
– Select the load type through the Ma-
terial selection key; the correspond-
ing LED switches on:
– TIMER display shows the exposure
time in minutes (20 / 30) of the se-
lected cycle.
Time (min.),
pre-vacuum
and heating
times excluded
35’ total =
20’ disinf.+
10’ drying +
5’ leveling
45’ total =
30’ disinf. +
10’ drying +
5’ leveling
30’ total =
30’ disinf.+
time for va-
pour conden-
sation
Optical fiber
handpeaces
Terminals of
polimerizing
lamps
Wrapped
optical fiber
handpeaces
Wrapped
terminals of
polimerizing
lamps
PARAMETERS OF THE STANDARD DISINFECTING CYCLES
CHOOSING A STANDARD DISINFECTING PROGRAM