INSTALLATION AND TRANSPORT
DIMENSIONS AND WEIGHT
Dimensions see figure
Weight 44 kg
REQUIRED VOLTAGE AND POWER
The voltage and power rating of the plant
where the autoclave is installed must comply
with the values reported on the rear identifica-
tion plate:
– voltage 230 Vac / 50 Hz
– current 8 A
– power 1700 W
– Position the autoclave on a plane
surface; adjust the rear feet to reach
a perfect leveling.
– Leave a free space of about 10 cm.
all around the sterilizer to allow
enough ventilation.
– Do not place the sterilizer near any
sink etc. in order to avoid any con-
tact with water and consequently
the short-circuit on the electrical cir-
cuits.
– Do not install the autoclave in envi-
ronment with inflammable gas, high
humidity or poor ventilation.
– The autoclave may be fitted in furni-
ture with an open space to enable
sufficient air circulation.
– Install the sterilizer so that the power
supply cable will run freely to the a.c.
outlet.