EXACTA
26
Switch on the autoclave by pushing on the right
side mains switch.
* THIS SWITCH CAN BE LEFT PERMANENTLY
IN ON POSITION AS THE MACHINE GOES
IN REST CONDITION AUTOMATICALLY
(AUTO SWITCH OFF), AT VERY LOW POW-
ER CONSUMPTION, AFTER 30 MINUTES
WITHOUT ANY ACTIVITY BY THE OPERA-
TOR.
The display Temperature indicates the Off state
of the autoclave.
Push on the ON/OFF button to enable the mi-
croprocessor functions.
Now the Temperature and Pressure displays
show as first the parameter values of the cycle
automatically selected at the switching on (e.g.
134°C - 2.10 bar) and then the current values in
the chamber.
The Timer display shows the time of the current
sterilization cycle.
In the first installation or whatever when the
distilled water is scarce in the tank, the red LED
of minimum level is flashing and the LED READY
remains off.
Fill the supplied tank with 2 liters of distilled
water, and keeping the tank lower plug the
supplied tube into the fast filling point; then
place the tank over the machine, unscrew
slightly its cap until the water starts to flow.
Reached the maximum level, the green LED
switches on and a beep sounds as confirmation.
* UNPLUG THE TUBE BY PUSHING ON THE
TOP LEVER OF THE FITTING AND DRAWING
THE TUBE. TAKE CARE TO BRING THE
TANK LOWER THE FILLING POINT LEVEL.