EXACTA
36
The selection may be carried out with door
open or close.
Fast cycle is a program at 134 °C/2,10 bar/04
min. without pre-vacuum phase and final drying
(DRY). ). This is a fast cycle suitable for loosing
tools that have to be used immediately.
* THIS CYCLE IS NOT SUITABLE FOR HOL-
LOW, WRAPPED OR POROUS MATERIALS
WHERE THE PRE-VACUUM PHASE IS A
VERY IMPORTANT CONDITION FOR THE
STERILIZATION.
Press DRY-OFF key and more times
the Temperature key until LED FAST
switches on. Release the keys;
LED 134°C switches on;
LED STER. switches on:
This function allows to program temperature
and time values that differ from the standard
ones. The time ranges from a minimum value,
depending on the selected temperature value,
up to a maximum of 99 minutes. The lowest
temperature and time for a sterilization cycle is
115°C / 30 min. At temperature lower than
115°C the cycle will be a disinfecting type.
Temperatures higher or equal to 115°C, the
sterilization or disinfecting cycles can be pro-
grammed with different times.
Press DRY-OFF key and more times
the Material selection key until LED
PROG switches on. Release the keys;
To increase the temperature value,
press the TEMPERATURE key, the dis-
played value increases a unit for
each step;
To increase the exposition value,
press the MATERIAL key, the dis-
played value increases a unit for
each step;
Verify which LED switches on (STER. or