– Place the tray-holder and the trays
with the load into the chamber.
– Choose the suitable cycle depend-
ing on the load and close the door.
– Push on the START key.
* THE CYCLE DOES NOR START IF LED
READY IS SWITCHED OFF. VERIFY THE
ALARM CODE ON THE TIMER DISPLAY AND
CORRECT THE PROBLEM.
EXAMPLE FOR THE CYCLE 134°C/2,1 bar
As the cycle starts, the LED READY flashes ini-
tially, then switches on steady after the first
distilled water filling phase (FILL1). The next
phases are signaled through proper codes on
the timer display.
Start of the thermodynamic vacuum phase.
Reached the setup value for the pre-heating
(about 80°C), the first pre-vacuum phase starts.
It follows a first distilled water filling phase.
Start of the heating phase.
Heating and filling phases alternate until the
values 134°C and 2,1 bar are reached into the
chamber.
It follows the high pressure phase and the stabi-
lization of the setup values.
Start of the sterilization phase with the count
down on the timer display.
Ended the count down, the cycle proceeds with
the pressure decreasing, count down of the
vacuum drying phase for 9 minutes, venting
and leveling phases for 2 minutes.