Expired the last count down and reached a
chamber pressure next to 0,0 bar, the tempera-
ture display will show the END message. If the
setup parameters have been agreed, LED STER
will flash to indicate the correct end for the
process.
Now it is possible to open the door and recov-
ery the sterilized load. As the door is opened
the front panel will fall back to the rest status
signaling.
The cycle may be interrupted at any time.
Push the START/STOP key for about 3 seconds
until the timer display shows the error code F3.
Depending on the phase interrupted, the ma-
chine enables a set of processes in order to
reach as soon as possible the opening condition
for the chamber.
Expired this time, the temperature display will
show the END message and the door can be
opened for the load recovery or for setting a
new cycle.
As the door is opened the front panel will fall
back to the rest status signaling.
After a set of cycles, the used water of the re-
covery tank reaches the maximum level sig-
naled on the timer display by a set of hyphens,
and the LED READY turns off.
LEAVE THE AUTOCLAVE IN REST CONDITION
FOR ABOUT 30 MINUTES IN ORDER TO ALLOW
THE USED WATER COOLING.
Plug the supplied tube into the fast fitting of
the draining point and e leave the water flow
into an external tank. For unplugging the tube,
push on the top lever of the fitting and draw
the tube.