Plug the tube into the fast fitting of the draining
point and empty completely the distilled water
tank; then fill up the supplied tank with 2 liters
of distilled water mixed with 100 ml of pure
alcohol (do not use denatured alcohol which
leaves residues).
Empty again the tank.
Ended this operation, fill up the tank with two
liters of distilled water to complete the supply (
until the LED MAX turns on).
For unplugging the tube, push on the upper
lever of the fitting and draw the tube.
Open the door and unscrew the screw (1) by
means of the supplied wrench. Remove the
dam (2) and the gasket.
By using the same wrench remove the fitting
(3) and the filter (4).
Clean the filter with water.
Plug the filter into the fitting (use a sealing type
Loctite, if possible); mount all the parts follow-
ing the previous procedure in the reverse order
and take care for keeping the holes (5) not low-
er of the chamber surface level.