EasyManua.ls Logo

MSA Chillgard RT - Cross-Sensitivity Response Data; Table 1-4. Typical Chillgard RT System for R-22 Cross-Sensitivity Response Data

MSA Chillgard RT
124 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
1-10
Ta
ble 1-3. Typical Chillgard RT System for
R-11 Cross-Sensitivity Response Data
GAS CONCENTRATION (PPM) EQUIVALENT PPM R-12
Acetone 100 3
Methyl Ethyl Ketone 1000 25
Methanol 1000 2
Methylene Chloride 1000 14
Trichloroethylene 1000 167
Ethyl Acetate 1000 22
Xylene (ortho-Xylene) 1000 5
R113 100 35
R11 100 5
R22 100 1
R12 100 100
R134A 100 2
R114 100 45
TEST CONDITIONS: Calibrated 0-1000 PPM R-12 in N
2;
Temperature - 25°C
Table 1-4. Typical Chillgard RT System for
R-22 Cross-Sensitivity Response Data
GAS CONCENTRATION (PPM) EQUIVALENT PPM R-22
Acetone 100 <1
R11 100 16
R123 100 35
R134A 100 25
R1132A 100 40
R22 100 100
R113 100 39
TEST CONDITIONS: Calibrated 0-1000 PPM R-22 in N
2
; Temperature: 25°C