OperationMSA
PrimaX
GB
25
(10) Set the idle time countdown in seconds with the UP or DOWN button.
(11) Press ENTER button.
[M-05] - Sensor Exchange
This menu initiates the initial calibration and resets the sensor lifetime counter.
OX/TOX sensors can be exchanged under power [hot swapped] with this function.
Attention!
Hot swaps must not be carried out for COMB sensors.
Only OX/TOX sensors can be exchanged with this menu.
For COMB sensors with this menu only the initial calibration is initiated
and the sensor life time is reset.
If an [OX/TOX] sensor is connected: If a COMB sensor is connected:
(1) Press ENTER and select
menu 5.
(2) Press ENTER button.
(3) Enter password [ chapter 4.5].
(1) Press ENTER and select
menu 5.
(2) Press ENTER button.
(3) Enter password [ chapter 4.5].
- An hourglass symbol is shown to
signify the time span [max. 15 min]
during which it is possible to hot-
swap a sensor without triggering
an error code
(4) [Perform a ZERO and SPAN
calibration [M-01].
(5) Change the sensor during this
time span.
(6) Press ENTER or ESC to start the
running-in time countdown.
(7) [Perform a ZERO and SPAN or
Oxygen calibration [M-01].
Sensors that are no longer required have to be disposed of in an envi-
ronmentally compatible way.