Operation
PrimaX
26
GB
[M-06] - LCD/LED Test
LCD and LED test [LED only available in PrimaX P]. All segments will be displayed
and the LEDs will flash sequentially.
(1) Press ENTER and select menu 6.
(2) Press ENTER button.
[M-07] - Loop Test
In this menu a 4 - 20 mA loop test will be performed.
(1) Press ENTER and select menu 7.
(2) Press ENTER button.
(3) Enter password [ chapter 4.5].
The loop test mA value is displayed [default value = 12 mA].
(4) Press UP or DOWN to change the value.
(5) Press ENTER button to start the test.
(6) Press ENTER button to go back to menu or ESC to go back to measuring
mode.
[M-08] - Device Information
In this menu device information is shown, i.e. gas type, measuring range, firmware
version.
Go through the information with the ENTER button.
[M-09] - Sensor Information
This menu shows the minimum and maximum measured gas concentrations, which
can be reset by holding the UP or DOWN button.
It shows the sensor life time in months since sensor exchange [ M-05].
If a COMB or a TOX sensor is connected this menu shows the response time [as
calculated during zero span calibration].
If an OX/TOX sensor is connected this menu shows the measured mV value.
If a COMB sensor is connected this menu shows the detector [U
D
], compensator
[U
K
] and differential voltage [U
X
] in mV.
Go through the information with the ENTER button.