Changeable Parameters for Relays
Changeable Parameters with HART
Values default minimum maximum
Alarm threshold 30% of Range
[O
2
= 20 vol%]
5 % of Range 100% of Range
Alarm relay normally en-
ergised
yes yes no
Latch alarm yes yes no
Alarming direction O
2
decrease;
other increase
Increasing Decreasing
Alarm delay time 0 s 0 s 600 s
Failure relay normally en-
ergised
yes yes no
Failure relay delay time 0 s 0 s 600 s
Setting required for ATEX conditions:
Latch alarm = yes: Alarm output self-retaining
Alarm delay time = 0s, no delay of output
Failure relay normally energised = yes
For all parameters which can be changed over keypad see chapter 4.6.
For relay parameters which can be changed over keypad see above.
In case of parameter modifications the functions have to be checked be-
fore normal operation.
Values default minimum maximum
Password 0000 0000 9999
Tag MSA - -
Description PrimaX - -
Long tag - -
Message - -
Enable LOC if combustible enable enable disable