Operation
PrimaX
32
GB
[M-12] - Relay Setup
(1) Press ENTER and select menu 12.
(2) Press ENTER button.
(3) Enter password [ chapter 4.5].
(4) Set the alarm threshold with the UP or DOWN button.
(5) Press ENTER button.
(6) Set rising or falling alarm option with the UP or DOWN button.
(7) Press ENTER button.
(8) Set if latched with the UP or DOWN button.
(9) Press ENTER button.
(10) Set the alarm relay delay with the UP or DOWN button.
(11) Press ENTER button.
(12) Set alarm relay normally energised option with the UP or DOWN button.
(13) Press ENTER button.
(14) Set the failure relay delay with the UP or DOWN button.
(15) Press ENTER button.
(16) Set the failure relay normally energised option with the UP or DOWN button.
(17) Press ENTER button.
[M-13] - Relay Test
(1) Press ENTER and select menu 13.
(2) Press ENTER button.
(3) Enter password [ chapter 4.5].
(4) The relays are now Switched and can be tested.
(5) Press ENTER button to go back to menu or ESC to go back to measuring
mode.