Gas
(Code)
Thread
Type
4
TruCal
Range
Default
Display
Resolution
3
Unit
Default
Alarm
1
Default
Alarm
1
Default
AO
(mA)
Alarm
2
Default
Alarm
2
Default
AO
(mA)
Alarm
Action
Default
Span
Value
Default
Alarm
Min
Alarm
Max
Range
Min
Range
Max
Span
Value
Min
Span
Value
Max
Sulfur
Dioxide
(51)
Coarse
0-100
PPM
0.1 PPM 10 5.6 20 5.28
Increasing
Non-Latch
10 2 100 0-10 0-100 7
FS
1
1
FS=FullScaleRange
2
TheRangeMaxvalueoncatalyticbeadcannotbesetbelow20%.
3
Displayresolutionisnotaconfigurableoption.
4
ClassIDivision2/Zone2Onlysensorsdonothaveaflamearrestor(akaFrit).Coursethreadsonthesensorassemblyandsensorbodyareusedtopreventacustomerfrom
installingintoaClassIDivision1/Zone1sensorbody.
TruCalnotavailable
AdaptiveEnvironmentalCompensation(AEC)Equipped
AECandDiffusionSupervisionEquipped