MTS 793 Controller Hardware FlexTest GT Controller Connections
143
How to connect
hydraulic terminals on
multiple station 407
Controllers
The following figure shows how to cable hydraulic terminals on
multiple station Model 407 Controllers, where each Model 407
Controller can run an independent test program. Repeat this scheme
for each additional Model 407 Controller.
Hyd
In
Hyd
Out
Intlk
In
Intlk
Out
407 Controller
No Connection
Jumper Plug PN
049-635-901
Jumper Plug PN
049-635-901
No Connection
Hyd
In
Hyd
Out
Intlk
In
Intlk
Out
407 Controller
No Connection
Jumper Plug PN
049-635-901
Jumper Plug PN
049-635-901
No Connection