MTS 793 Controller Hardware
FlexTest GT Controller Connections
144
Interlock connections
for master/dependent
407 Controllers
The following figure shows how to cable interlocks when a master
Model 407 Controller is daisy-chained with three other dependent
Model 407 Controllers to run a single test program on a FlexTest GT
Controller (via its Model 493.74 HSM transition module). Use the
specified four-channel master/dependent interlock cable (PN 056-448-
9XX) to make these connections for each Model 407 Controller station.
Jumper
Jumper
Jumper
Digital
I/O Connector
Hydraulic
Interlock
Active Low
Hydraulic
Interlock
Active Low
Hydraulic
Interlock
Active Low
Hydraulic
Interlock
Active Low
Run/Stop
Active High
Run/Stop
Active High
Run/Stop
Active High
To
407 Controller
To
407 Controller
To
407 Controller
To
407 Controller
Run/Stop
Active High
To
493.74
HSM
Transition
Master/Dependent Interlock Cable
(PN 056-448-9XX)
Master
Dependent
Dependent