EasyManua.ls Logo

Nicoya OpenSPR - Page 53

Nicoya OpenSPR
61 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
53
OpenSPR 1-Channel Operation Manual V4.1.01
Figure 2.72 Shutdown process 2/3: Sterilization with 80% isopropanol.
5. A progress bar will display the sterilization step status. During this time, rinse the injection
port with at least 5 mL of 80% isopropanol.
6. The third step in the shutdown process will be to purge the pumps and tubing of the
system with air [Figure 2.73]. To do this, remove all 3 inlet tubes from solution and
suspend the ends of the tubes in a clean, safe space. Once all inlet tubes are in proper
position, click “Next” to begin the purging step.
Figure 2.73 Shutdown process 3/3: purging with air.
7. A progress bar will display the status of the purging step. During this time, purge the
injection port with at least 2 mL of air.
8. Once the shutdown preparation processes are complete, a final dialogue box will appear
[Figure 2.74]. Clicking “OK” in this dialogue will safely close the software and power off
the OpenSPR instrument automatically.
Figure 2.74 Shutdown process complete.

Other manuals for Nicoya OpenSPR

Related product manuals