Chapter 2 Microscopy
2.5 Photomicroscopy
26
2.5
Photomicroscopy
For detailed discussions of the camera, photomicroscopic software, and PC, refer to the
operating manuals provided with the respective products. The following instructions assume a
DS camera head DS-5M and DS-L1 camera control unit.
1
Adjust the microscope for proper image observation.
See the directions given in sections from “2.1 Diascopic Bright-field Microscopy” to “2.4
Episcopic Microscopy.”
2
Adjust the camera head attaching position until the image is
displayed properly.
Loosen the attachment guide fixing screw on the C
mount and adjust the camera position so that
moving the stage left-right moves the image on the
monitor in the opposite direction. After making the
appropriate adjustments, tighten the screws firmly.
Adjust the directions (A) and (B).
3
Make camera settings.
For detailed information, refer to the operating manual provided with the camera.
When using the DS-L1, you must choose and enter at least the following information:
Folder for data storage
Name of file to be saved (You can select “Auto.”)
File format and file size
Date and destination of data
4
Select the camera scene mode suitable for the microscopy
method.
5
Set the camera white balance.
To adjust the white balance, press the WB button while capturing an image of a clear section of
a specimen slide. For fluorescent photomicrography, adjust the white balance under normal
lighting conditions before the image capturing.
(A)
(B)