62
Replacing Consumables
5
5.1
Replacing the lamp
Be careful to avoid burns:
Wait until the lamp and nearby parts have cooled before attempting to
replace the lamp.
Be careful to avoid electrical shock:
Turn off the power switch and unplug the power cord from the outlet.
Be careful to avoid abnormal heat generation:
Use only the lamp specified.
Be careful to avoid soiling:
Avoid touching the bare lamp bulb with bare hands. Soiling will reduce the
service life of the lamp.
Replacing the lamp for the diascopic illumination
(1) Remove the lamphouse cover on the back of
the microscope. Remove the old lamp.
(2) Attach a new lamp.
Avoid touching the bare lamp bulb with your
bare hands.
(3) Replace the cover.
Engage the cover hook in the slot on the rear
of the unit in the direction indicated by the
arrow in the figure.
Be sure to attach the lamphouse cover. Failure to do so may result in burns or
fire from the heat generated by the lamp.
CAUTION
CAUTION
Lamp
Lamphouse cover
Cover hook
Use only the lamp specified (PHILIPS
5761).
Slot for cover hook