Chapter 4 Assembly
65
4
6
Installing a Fluorescent Attachment (when using a
fluorescent attachment)
(1) Place a fluorescent attachment on the
microscope arm.
(2) Secure in place with a clamp screw.
(3) Fix the fluorescent attachment clamp bolt
using the hex wrench provided with the unit.
(4) Attach a mercury lamphouse to the bayonet
mount on the back. (For more information,
refer to the instruction manual provided with
the superhigh pressure mercury lamp power
supply.)
1) Attach a collector lens to the lamphouse.
2) Turn the bayonet mount clockwise as far as
it will go.
3) Insert the lamphouse into position.
4) Turn the bayonet mount back to its original
position to lock the lamphouse.
5) Attach a mercury lamp.
6) Connect the mercury lamphouse to the
mercury lamp power supply.
7
Installing an Eyepiece Tube
(1) Place an eyepiece tube on the microscope arm (or fluorescent unit).
(2) Secure in place with a clamp screw. (For the fluorescent unit, secure in place with a fixing
screw using the tool stored in the back of the microscope.)
Clamp screws used to fix
the fluorescent unit
Ba