Attaching the Circular Graduated Stage
1 Remove the cushioning material from the
substage section and turn the coarse focus
handle until the elevating section is brought
to the uppermost position.
2 Place the circular graduated stage on the
substage. The stage must be level and both
stage rotation clamp screws must be located
in 45° directions from the microscope front
when fixed.
3 Fix the circular graduated stage with four
hexagonal socket head bolts via the sub-
stage in the upward direction. Use the
hexagonal wrench (2.5 mm) provided with
the microscope to screw the bolts.
4 If necessary, attach the specimen holders
into the hole on the stage.
For information about the optional
attachable mechanical stage, see Page 26,
“Attachable Mechanical Stage (Option).”
2
Attaching the Condenser
To perform the polarization microscopy, the P swing-out condenser must be attached.
1 Turn the coarse focus knob until the
elevating section is raised to the uppermost
position.
2 Turn the condenser focus knob until the
condenser holder is brought to the
lowermost position.
3 Insert the condenser and adjust its
orientation so that it faces toward the front.
Secure in place using the tool provided with
the microscope.
4 Turn the condenser focus knob until the
condenser holder is raised to the uppermost
position.
ND8
N
CB
F
.S.
Substage
Hexagonal
socket head
bolts (x4)
Circular
graduated
stage
EPI/DIA
OFF
0.6
0.8
0.
2
0.
4
J
APAN
A
chr
0.
90
CLICK
ON
Condenser
focus knob
Condenser
Condenser
clamp screw
Coarse
focus knob
0.6
0.8
0.
2
0.
4
J
APA
N
Achr
0.
90