62
Specifications
Model
Optical system
Illumination system
Focusing mechanism
Eyepiece
Stage
Nosepiece
Input ratings
Power cord
Operating conditions
Storage conditions
ECLIPSE LV100POL
CFI60 system (chromatic aberration free infinity system)
Epi/dia selector type
Episcopic illumination: Built-in type lamp power supply
Specified illuminator: LV-UEPI Universal Epi Illuminator (NCB11, ND4, and ND16
are installed and can be replaced by users)
Diascopic illumination: Built-in type lamp power supply, fly's eye lens, NCB11 and
ND8 are installed but cannot be replaced by users
Lamp ratings: 12 VDC, 50 W halogen lamp
Specified lamp: LV-HL50W, 12 V, 50 W longlife halogen lamp
Specified lamp house: LV-LH50PC precentered lamp house
Manual operation type single axis coarse/fine focus knob mechanism (calibration marking
for fine focus: 1 µm/marking)
Stroke: 30 mm with coarse focus stopper mechanism
Coarse focus knob: 14 mm/revolution
Fine focus knob: 0.1 mm/revolution
10x, field number: 22
Circular graduated stage
Provided with two verniers
Provided with a rotation mechanism with a clamp
Provided with 45° click-stop mechanism
Provided with two specimen holders
Quintuple type (five sockets)
Provided with objective centering mechanism
Input voltage: 100-240 VAC ±10% 50/60 Hz
Rated current: 1.2 A maximum
When the supply voltage is 100 V to 120 V
UL Listed detachable cord set, 3 conductor grounding Type
(3 conductor grounding Type SVT, No.18 AWG, 3m long
maximum, rated at 125V AC minimum)
When the supply voltage is 220 V to 240 V
Approved according to EU/EN standards, 3 conductor
grounding Type (3 conductor grounding Type H05VV-F, 3m
long maximum, rated at 250V AC minimum)
Temperature: 0°C to +40°C
Humidity: 85% relative humidity maximum (no condensation)
Altitude: 2000 m maximum
Pollution degree: Degree 2
Installation category: Category II
Electric shock protection class: Class 1
Indoor use only
Temperature: -20°C to +60°C
Humidity: 90% relative humidity maximum (no condensation)