5.2 Functional sequence
If the toggle switch is turned to the ‘On‘ position, the magnetic valve (1) opens and water flows
into the system. After a short delay, the pump (4) will start.
Now the water is pressed into the reverse osmosis membrane at a pressure of 10-15 bar. The
flow is divided into a permeate and a concentrate part. The permeate content flows through the
flow indicator (7), the temperature sensor (8), the conductivity sensor (9) and the pressure
switch (10) into the ring line. The unused permeate flows back into the RO Medical-Basic via the
connection (12).
To save water, the concentrate portion is divided up again, one portion is given into the drain via
the flow indicator (16), the other is fed back in front of the pump.