12 | 04.10.2021 | Page 15
6.3 Electrical installation
The installation may only be performed by a qualified electrician.
The device must be supplied by a permanent connection, connectors are not
valid. The disconnection via the main switch at the control cabinet. The power
cord must be provided with a strain relief.
Connection cable RO Medical-Basic
For protection against a re-start of the unit, the main switch
can be locked with a padlock.
Safety class I
The device is equipped with a Protective earth terminal for prevention against
high touch current
For prevention of the hazard of an electric shock, this device may only be
connected to a power supply with protective earth.
The power cord is fixed to the device and cannot be replaced.