7.3 Device on / off with permeate tank
If the RO Medical-Basic is connected to a permeate tank, the toggle switch
must be set to "Auto". Since the device switches off automatically when the
tank is full, it is not necessary to reset it to position 0.
Caution, danger of overflow!
If the RO Medical-Basic is operated with a permeate tank, the system may only
be started via "Auto". In the "On" position there is a risk of overflow.
Device on
To start the system, turn the knob switch to the right.
If the permeate tank is empty, the system starts automatically.
If the tank is already full, the osmosis waits to start until the level
in the tank drops.