12. Service Parameters
The settings and functions described below may only be carried out by technically
trained personnel.
ATTENTION, SYSTEM DAMAGE!
Incorrect settings can lead to damage.
Arrow key down and then press the ESC key. (Service mode)
Then Logo settings> ENTER> Program> ENTER> Set
parameters> ENTER> select the appropriate parameter (B001-
B040)
Parameter Function
default value
T Pump
Pump start time delay
3sec
Time to Rinse
Rinse waiting time
60m
T Rinse
Rinse time
10m
T
Conductivity
Conductivity alarm delay at start
2m
W Qty NOK
Dry running protection
5sec
T2 Conductivity
Conductivity alarm delay at operation
30sec
B025
Hours count
T Temperatur
Temperature alarm delay
10sec
Conductivity
Conductivity adjustment (permeate)
B040
Timer (auto start)