2. Intended operation
The unit is designed for the treatment of potable water. The pure water (permeate) thus
produced may be used for dialysis treatment.
Other applications are only possible after consulting the manufacturer and receiving their
approval.
The unit can only be maintained by the manufacturer or technicians trained by
the manufacturer.
Only original replacement parts may be used for maintenance and repairs.
Installation operations, modifications or reparations, are only allowed to be
performed by persons authorised by the manufacturer and may only be done
with original replacement parts. Improper performed reparations or modifications
The device may only be operated in perfect condition.
Before operating, check the following:
Lose or defect parts
Defect cables and/or isolations
Serious soiling
The device may only be operated with the appropriate ring line.
The device does not produce water for injections.
The device has pressurized parts.
If the temperature sensor fails, the temperature in the permeate can increase.
(Max 60°C)
The water treatment system RO Medical-Basic may only be used for permeate
supply of dialysis devices, which have a temperature measurement (permeate
temperature).
The device has no direct patient contact and no patient application part.
2.1 Contraindications / side effects
None