10.2 Change the conductivity parameter
Press the F4 key and F2 key simultaneously.
The conductivity in the return is displayed.
Press the ESC key for 3 seconds, the first value is marked.
Press enter
Set the value with the arrow keys (up / down). Complete the
entry with ENTER.
Press the ESC key.
Press the F4 and F2 keys simultaneously to return to the
standard display
Both parameters (MAX1 and MAX2) must be set to the same value.
Standard value = 100 µS / cm
If the conductivity in the back flow reaches the value set here, the conductivity
alarm is activated.