11. Replacement of the reverse osmosis
membrane
Caution pressure!
Membrane tubes are under pressure. Please open carefully.
The settings and functions described below may only be carried out by technically
trained personnel.
1. Switch off the system at the main switch.
2. Open wing screw and take off the clip
3. Lift module cover by using a plier
4. Take off cover.
5. Pull out the membrane by using a plier
6. Reinstall the new membrane in reverse order.
Take care of the flow direction and
position of the gasket
Rinse membrane!
After the new membrane has been installed the mode “drain permeate”
has to be started for 20 minutes.